......@@ -57,7 +57,17 @@ The rocker switch ③ can be used to select between internal manual unblanking (
### Test of the supply current
By measuring the supply current I_SC, the general function of the individual amplifiers (High Power Amplifier ZHL-20W-13SW+) is to be tested. For this purpose, a multimeter (VOLTCRAFT MT-51, SN: 09126841) was inserted into the current path of the supply current I_SC. According to the amplifier's data sheet, the supply current I_SC is max. 2.8 A during proper operation (with fan). The values measured (Table 2) are within the tolerances.
(table 2 here)
| Channel No. | ΔU<sub>out</sub> @f=123 MHz P<sub>in</sub>: | Function | ΔU<sub>out</sub> @f=123 MHz P<sub>in</sub>: | |
| ----- | -------- | -------- | ----- | ----- |
| 1 | SMA connector | RF input | XX | 8 |
| 2 | BNC connector | Input external 5V TTL signal | XX | 1 |
| 3 | Toggle switch | TTL or manual blanking | XX | 1 |
| 4 | On/off-switch | 230 V power switch | XX | 1 |
| 5 | SMA connector | Output RF signal | XX | 16 |
| 6 | Fuse | 10 A / electrical protection | XX | 1 |
| 7 | C14 connector | Main power supply connector | XX | 1 |
| 8 | 24V Power supply | RSP-1000-24 | XX | 1 |